- A Spectrophotometer Was Calibrated With A 7 300 X 10 4 M In Cu Cu Neocuproine Solution This Calibration Was Carried O 1 (69.18 KiB) Viewed 57 times
A spectrophotometer was calibrated with a 7.300 x 10-4 M (in Cu) Cu-neocuproine solution. This calibration was carried o
-
- Site Admin
- Posts: 899603
- Joined: Mon Aug 02, 2021 8:13 am
A spectrophotometer was calibrated with a 7.300 x 10-4 M (in Cu) Cu-neocuproine solution. This calibration was carried o
A spectrophotometer was calibrated with a 7.300 x 10-4 M (in Cu) Cu-neocuproine solution. This calibration was carried out in a 1.00 cm cell, and yielded an absorbance of 0.2920 AU. An unknown was then analyzed in this same spectrophotometer, but using a 2.50 cm cell. If the absorbance of the unknown was 0.4020 AU, what is the Cu concentration? 4.0 0.000402 X M