c) What is the maximum safe value of the R_ground in the below schematic so that there appear no microshock hazard? Calc
-
answerhappygod
- Site Admin
- Posts: 899604
- Joined: Mon Aug 02, 2021 8:13 am
c) What is the maximum safe value of the R_ground in the below schematic so that there appear no microshock hazard? Calc
c) What is the maximum safe value of the R_ground in the below schematic so that there appear no microshock hazard? Calculate it. hot R_device 10k SINE(0 311 50) R_body 500 neutral C R_ground 400p .tran 0 0.1 0 0.000001 NOTE: The maximum current allowed through the heart of the patient is 10 μArms. V1